48
3.2 Trigger Operations
The triggering system of the DL-2060 will be explained in this section.
DL-2060 provides a various trigger operations such as trigger mode, trigger source,
and different trigger settings for a specific measurement. The selected setup will be
stored into the volatile memory and it returns to the default mode after power off.
Following
Figure 4-8
describes the detail of the trigger setup.
Figure 4-8
The trigger system is supposed to be in an idle status when DL-2060 is not measuring.
( *(sampling)display will illuminates when measurement is in progress )
When starting measurement after idle status, the trigger system should be in trigger
waiting status. To make is in trigger waiting status, execute one of the following
command via remote interface.
INTiate
READ?
MEASure?
While DL-2060 is in trigger waiting status(Waiting for event), the measurement operation
will starts after execution of trigger delay, when the trigger condition matches by
detecting trigger signal from the selected trigger source etc. The operation form the front
panel will always be trigger waiting status then receives trigger signal at any time.