Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-20
2470-901-01 Rev. A /
May
2019
The results of the limit test for limit
Y
:
•
NONE
: Test passed; the measurement is between the upper and lower limits
•
HIGH
: Test failed; the measurement exceeded the upper limit
•
LOW
: Test failed; the measurement exceeded the lower limit
•
BOTH
: Test failed; the measurement exceeded both limits
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage measurement
exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results display on the
front panel.
Return the test results; example output if the test fails on the
low limit:
LOW
Clear the test results.
Also see
:CALCulate2:<function>:LIMit<Y>:CLEar:AUTO
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
(on page 12-18)
:CALCulate2:<function>:LIMit<Y>:STATe
(on page 12-22)
(on page 4-66)
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
This command specifies the lower limit for limit tests.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
-1
Usage
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA] <n>
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA] <DEF|MIN|MAX>
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]?
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<Y>
Limit number:
1
or
2
<n>
The low limit value of limit
Y
(-9.11 to 9.11)
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value