Section 1: Introduction
2470 High Voltage SourceMeter Instrument Reference Manual
1-4
2470-901-01 Rev. A /
May
2019
General ratings
The 2470 instrument's general ratings and connections are listed in the following table.
Category
Specification
Supply voltage range
100 V
RMS
to 240 V
RMS
, 50 Hz or 60 Hz (automatically detected at power
up)
Input and output connections
See
Environmental conditions
For indoor use only
Altitude:
Maximum 2000 meters (6562 feet) above sea level
Operating:
0 °C to 50 °C, 70% relative humidity up to 35 °C; derate 3%
relative humidity per °C, 35 °C to 50 °C
Storage:
−
25 °C to 65 °C
Pollution category:
2