Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-164
2470-901-01 Rev. A /
May
2019
:TRIGger:BLOCk:LIST?
This command returns the settings for all trigger model blocks.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:BLOCk:LIST?
Details
This returns the settings for the trigger model.
Example
:TRIG:BLOC:LIST?
Returns the settings for the trigger model. Example output is:
1) BUFFER_CLEAR
BUFFER: defbuffer1
2) MEASURE
BUFFER: defbuffer1 COUNT: 1
3) BRANCH_COUNTER
VALUE: 5 BRANCH_BLOCK: 2
4) DELAY_CONSTANT
DELAY: 1.000000000
5) BRANCH_COUNTER
VALUE: 3 BRANCH_BLOCK: 2
Also see
None
:TRIGger:BLOCk:LOG:EVENt
This command allows you to log an event in the event log when the trigger model is running.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:LOG:EVENt <blockNumber>, <eventNumber>, "<message>"
<blockNumber>
The sequence of the block in the trigger model
<eventNumber>
The event number:
INFO<n>
WARNing<n>
ERRor<n>
Where
<n>
is 1 to 4; you can define up to four of each type
You can also set
ABORt
, which aborts the trigger model immediately and posts a
warning event log message
<message>
A string up to 31 characters