2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-139
smu.measure.nplc
This command sets the time that the input signal is measured for the selected function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
1
Usage
nplc
= smu.measure.nplc
smu.measure.nplc =
nplc
nplc
The number of power line cycles: 0.01 to 10
Details
This command sets the amount of time that the input signal is measured.
The amount of time is specified as the number of power line cycles (NPLCs). Each PLC for 60 Hz is
16.67 ms (1/60) and each PLC for 50 Hz is 20 ms (1/50). For 60 Hz, if you set the NPLC to 0.1, the
measure time is 1.667 ms.
This command is set for the measurement of specific functions (current, resistance, or voltage).
The shortest amount of time results in the fastest reading rate, but increases the reading noise and
decreases the number of usable digits.
The longest amount of time provides the lowest reading noise and more usable digits, but has the
slowest reading rate.
Settings between the fastest and slowest number of PLCs are a compromise between speed and
noise.
If you change the PLCs, you may want to adjust the displayed digits to reflect the change in usable
digits.
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.nplc = 0.5
Set the measurement function to DC Voltage.
Set the NPLC value to 0.5, which is 0.0083 seconds
(0.5/60).