Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-260
2470-901-01 Rev. A /
May
2019
Details
Set this command to any trigger event to start the timer when that event occurs.
Set this command to none to disable event processing and use the timer as a timer or alarm based
on the start time.
Trigger events are described in the table below.
Trigger events
Event description
Event constant
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
A command interface trigger:
Any remote interface: *
TRG
GPIB only: GET bus command
USB only: A USBTMC TRIGGER message
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Digital input line edge (either rising, falling, or either based on
the configuration of the line) detected on digital input line
N
(1 to 6)
trigger.EVENT_DIGIO
N
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Appropriate LXI trigger packet is received on LAN trigger
object
N
(1 to 8)
trigger.EVENT_LAN
N
No trigger event
trigger.EVENT_NONE
Notify trigger block
N
(1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFY
N
Source limit condition occurs
trigger.EVENT_SOURCE_LIMIT
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Line edge detected on TSP-Link synchronization line
N
(1 to 3)
trigger.EVENT_TSPLINK
N
Example
digio.line[3].mode = digio.MODE_TRIGGER_IN
trigger.timer[1].delay = 3e-3
trigger.timer[1].start.stimulus = trigger.EVENT_DIGIO3
Set digital I/O line 3 to be a trigger input.
Set timer 1 to delay for 3 ms.
Set timer 1 to start the timer when an event is detected on digital I/O line 3.
Also see
None