2470 High Voltage SourceMeter Instrument Reference Manual
Section 4: Sourcing and measuring
2470-901-01 Rev. A /
May
2019
4-9
Figure 57: Model 2470 2-wire rear-panel connections
Using the front panel:
1.
Press the
FUNCTION
key.
2.
Select the source and measure combination.
3.
Press the
MENU
key.
4.
Under Measure, select
Settings
.
5.
Select the button next to Sense and select
2-Wire Sense
.
6.
Press the
HOME
key to return to the operating display.
Using SCPI commands:
To change to 2-wire sensing for current measurements, send the command:
:SENSe:CURRent:RSENse OFF
To change to 2-wire sensing for voltage, replace
CURRent
with
VOLTage
. For resistance, replace
CURRent
with
RESistance
.
Using TSP commands:
For voltage measurements, send the commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.sense = smu.SENSE_2WIRE
To assign a different measure function, replace
smu.FUNC_DC_VOLTAGE
with one of the following:
•
For current measurements:
smu.FUNC_DC_CURRENT
•
For resistance measurements:
smu.FUNC_RESISTANCE