Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-196
2470-901-01 Rev. A /
May
2019
Details
This command indicates if an event was ignored because the event detector was already in the
detected state when the event occurred.
This is an indication of the state of the event detector built into the timer itself. It does not indicate if
an overrun occurred in any other part of the trigger model or in any other construct that is monitoring
the delay completion event. It also is not an indication of a delay overrun.
This returns
0
if there is no overrun or
1
if there is an overrun.
Example
TRIG:TIM1:STAR:OVER?
Checks the overrun status on trigger timer 1.
Also see
None
:TRIGger:TIMer<n>:STARt:SEConds
This command configures an alarm or a time in the future when the timer will start.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0
Usage
:TRIGger:TIMer<n>:STARt:SEConds <time>
:TRIGger:TIMer<n>:STARt:SEConds?
<n>
Trigger timer number (1 to 4)
<time>
The time: 0 to 2,147,483,647 s
Details
This command configures the alarm of the timer.
When the timer is enabled, the timer starts immediately if the timer is configured for a start time that
has passed.
Example
TRIG:TIM1:STAR:SEC 60
TRIG:TIM1:STAR:FRAC 0.5
TRIG:TIM1:STAT ON
Set the timer for 60.5 s.
Enable the trigger timer for timer 1.
Also see
(on page 12-198)