Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-266
2470-901-01 Rev. A /
May
2019
trigger.tsplinkout[N].pulsewidth
This attribute sets the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
TSP-Link line
N
reset
Configuration script
10e-6 (10
µ
s)
Usage
width
= trigger.tsplinkout[
N
].pulsewidth
trigger.tsplinkout[
N
].pulsewidth =
width
width
The pulse width (0.0 to 100 ks)
N
The trigger line (1 to 3)
Details
Setting the pulse width to 0 asserts the trigger indefinitely.
Example
tsplink.line[3].mode = tsplink.MODE_TRIGGER_OPEN_DRAIN
trigger.tsplinkout[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to
20
μs.
Also see
trigger.tsplinkout[N].assert()
(on page 14-264)
trigger.tsplinkout[N].release()
(on page 14-266)
(on page 14-271)
trigger.tsplinkout[N].release()
This function releases a latched trigger on the given TSP-Link trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.tsplinkout[
N
].release()
N
The trigger line (1 to 3)
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger.