Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-216
2470-901-01 Rev. A /
May
2019
Example
reset()
trigger.model.setblock(1, trigger.BLOCK_BUFFER_CLEAR)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(3, trigger.BLOCK_DELAY_CONSTANT, 0.1)
trigger.model.setblock(4, trigger.BLOCK_BRANCH_COUNTER, 10, 2)
trigger.model.initiate()
delay(1)
print(trigger.model.getbranchcount(4))
waitcomplete()
Reset trigger model settings.
Clear
defbuffer1
at the beginning of the trigger model.
Loop and make five readings.
Delay 0.1 s.
Loop ten more times back to block 2.
Send the count command to check which count has been completed for block 4.
At end of execution, 10 readings are stored in
defbuffer1
.
Also see
trigger.model.setblock() — trigger.BLOCK_BRANCH_COUNTER
(on page 14-227)
trigger.model.initiate()
This function starts the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.model.initiate()
Also see
(on page 8-27)