2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-137
:TRACe:STATistics:STDDev?
This command returns the standard deviation of all readings in the buffer.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:STATistics:STDDev?
:TRACe:STATistics:STDDev? "<bufferName>"
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Example
TRACe:STAT:STDDev?
Returns the standard deviation of the readings in the default
buffer
defbuffer1
.
TRACe:STAT:STDDev? "testData"
Returns the standard deviation of the readings in the
user-defined buffer
testData
.
Also see
(on page 6-25)
(on page 12-127)
(on page 12-134)
(on page 12-135)
(on page 12-135)
(on page 12-136)
:TRACe:TRIGger
This command makes readings using the active measure function and stores them in a reading buffer.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:TRIGger
:TRACe:TRIGger "<bufferName>"
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Details
A measure function must be selected before sending this command.
This command makes the number of measurements that is set by the count command.