Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-70
2470-901-01 Rev. A /
May
2019
display.waitevent()
This function causes the instrument to wait for a user to respond to a prompt that was created with a prompt
command.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
objectID
,
subID
= display.waitevent()
objectID
,
subID
= display.waitevent(
timeout
)
objectID
A number that identifies the object, such as a prompt message, that is displayed on
the front panel
subID
The returned value after a button is pressed on the front panel:
display.BUTTON_YES
display.BUTTON_NO
display.BUTTON_OK
display.BUTTON_CANCEL
timeout
The amount of time to wait before timing out; time is 0 to 300 s, where the default of
0 waits indefinitely
Details
This command waits until a user responds to a front-panel prompt that was created with the
display.prompt()
command.
Example
smu.source.sweeplinear("test", 1, 10, 10)
display.prompt(display.BUTTONS_YESNO, "Would you like to start the sweep now?")
sweepTest, result = display.waitevent()
if result == display.BUTTON_YES then
trigger.model.initiate()
end
display.prompt(display.BUTTONS_YESNO, "Would you like to switch to the Graph screen?")
promptID, result = display.waitevent()
if result == display.BUTTON_YES then
display.changescreen(display.SCREEN_GRAPH)
end
Create a linear sweep.
Display the prompt "Would you like to start the sweep now?"
If the user presses Yes, the sweep starts.
If the user presses No, the sweep does not start and the message is removed.
Display the prompt "Would you like to switch to the Graph screen?"
If the user presses Yes, the Graph screen is displayed.
If the user presses No, the user remains on the present screen.
Also see
(on page 14-63)
(on page 14-67)