2470 High Voltage SourceMeter Instrument Reference Manual
Section 4: Sourcing and measuring
2470-901-01 Rev. A /
May
2019
4-51
Reciprocal (1/X)
You can set math operation to reciprocal to display the reciprocal of a reading.
The reciprocal is 1/X, where X is the reading. If relative offset is on, the 1/X calculation uses the input
signal with the relative offset applied.
Example:
Assume the normal displayed reading is 002.5000 Ω. The reciprocal of resistance is conductance.
When the reciprocal math function is enabled, the following conductance reading is displayed:
0.400000
When the reciprocal math operation is active, the unit of measure for the front-panel readings is 1/x
and the MATH indicator is displayed to the right of the measurement. You cannot change this units
designator.
Setting percent math operations
From the front panel:
1.
Press the
FUNCTION
key and select the measure function.
2.
Press the
MENU
key.
3.
Under Measure, select
Calculations
.
4.
Set Math to
On
.
5.
Select
Config
.
6.
For Math Format, select
Percent
.
7.
For Zero Reference, select the percent reference.
8.
Select
OK
.
9.
Press the
HOME
key to view the measurement with the percent math format applied.
Over a remote interface:
•
:CALCulate[1]:<function>:MATH:FORMat
(on page 12-10)
:CALCulate[1]:<function>:MATH:PERCent
(on page 12-14).
•
(on page 14-135) and
(on page 14-138).