Section 6: Reading buffers
2470 High Voltage SourceMeter Instrument Reference Manual
6-22
2470-901-01 Rev. A /
May
2019
Using the front panel to save buffer content to files:
1.
Insert a USB flash drive into the USB port.
2.
Press the
MENU
key.
3.
Under Measure, select
Reading Buffers
. The MANAGE READING BUFFERS window is
displayed.
4.
Select the reading buffer that you want to save. For example, select
testData
.
Figure 120: MANAGE READING BUFFERS window
5.
Select
Save To USB
.
6.
A keyboard is displayed.
7.
Enter the name of the file in which to save the readings.
You only need to enter the name of the file you want to save. It is not necessary to enter the file
extension. All files are saved as CSV files.
8.
Select
OK
on the keyboard.
9.
Select
OK
to save the file. When the MANAGE READING BUFFERS window is displayed again,
the file is saved.
10. Press the
HOME
key to return to the home screen.