Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-14
2470-901-01 Rev. A /
May
2019
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STAT ON
Set the math function for voltage measurements to mx+b.
Set the scale factor for voltage measurements to 0.80.
Set the offset factor to 50.
Enable the math function.
Also see
Calculations that you can apply to measurements
(on page 4-49)
:CALCulate[1]:<function>:MATH:FORMat
(on page 12-10)
:CALCulate[1]:<function>:MATH:MBFactor
(on page 12-11)
:CALCulate[1]:<function>:MATH:STATe
(on page 12-15)
:CALCulate[1]:<function>:MATH:PERCent
This command specifies the reference constant that is used when math operations are set to percent.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1
Usage
:CALCulate[1]:<function>:MATH:PERCent <value>
:CALCulate[1]:<function>:MATH:PERCent <DEF|MIN|MAX>
:CALCulate[1]:<function>:MATH:PERCent?
:CALCulate[1]:<function>:MATH:PERCent? <DEF|MIN|MAX>
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<value>
The reference used when the math operation is set to percent; the range is -1e12 to
+1e12
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value
Details
This is the constant that is used when the math operation is set to percent.
The percent math function displays measurements as percent deviation from a specified reference
constant. The percent calculation is:
Where:
•
Percent
is the result
•
Input
is the measurement (if relative offset is being used, this is the relative offset value)
•
Reference
is the user-specified constant