2470 High Voltage SourceMeter Instrument Reference Manual
Appendix B: Status model
2470-901-01 Rev. A /
May
2019
B-3
Standard Event Register
The Standard Event Register set includes two 8-bit registers:
•
Standard Event Status Register:
Reports when a predefined event has occurred. The register
latches the event and the corresponding bit remains set until it is cleared by a read.
•
Standard Event Status Enable Register:
You can enable or disable bits in this register. This
allows the predefined event (from the Standard Event Status Register) to set the ESB of the
Status Byte Register.
Figure 162: 2470 Standard Event Register
Bit When set, indicates the following has occurred:
0
Operation complete: All pending selected instrument operations are complete and the instrument is
ready to accept new commands. The bit is set in response to an
(on page A-6) command or TSP
(on page 14-98) function.
1
Not used.
2
Query error: Attempt to read data from an empty Output Queue.
3
Not used.
4
Not used.
5
Not used.
6
Not used.
7
Power-on: The instrument has been turned off and turned back on since the last time this register was
read.