Section 8: Triggering
2470 High Voltage SourceMeter Instrument Reference Manual
8-54
2470-901-01 Rev. A /
May
2019
Trigger events — SCPI command set
Trigger events
Event description
Event constant
Trigger event blender
<n>
(up to two), which combines trigger
events
BLENder<n>
A command interface trigger:
Any remote interface: *
TRG
GPIB only: GET bus command
USB only: A USBTMC TRIGGER message
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Front-panel TRIGGER key press
DISPlay
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
No trigger event
NONE
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
Source limit condition occurs
SLIMit
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>
Trigger events – TSP command set
Trigger events
Event description
Event constant
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
A command interface trigger:
Any remote interface: *
TRG
GPIB only: GET bus command
USB only: A USBTMC TRIGGER message
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Digital input line edge (either rising, falling, or either based on
the configuration of the line) detected on digital input line
N
(1 to 6)
trigger.EVENT_DIGIO
N
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Appropriate LXI trigger packet is received on LAN trigger
object
N
(1 to 8)
trigger.EVENT_LAN
N
No trigger event
trigger.EVENT_NONE
Notify trigger block
N
(1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFY
N
Source limit condition occurs
trigger.EVENT_SOURCE_LIMIT
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Line edge detected on TSP-Link synchronization line
N
(1 to 3)
trigger.EVENT_TSPLINK
N