Example
:SENSe:CONF:LIST:RECall "MyMeasList", 5
Recalls configuration index 5 in a configuration list
named
MyMeasList
.
:SENSe:CONF:LIST:RECall "MyMeasList"
Because an index was not specified, this command
recalls configuration index 1 from a configuration list
named
MyMeasList
.
Also see
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 12-61)
(on page 12-9)
[:SENSe[1]]:CONFiguration:LIST:STORe
[:SENSe[1]]:CONFiguration:LIST:SIZE?
This command returns the size (number of configuration indexes) of a measure configuration list.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:SIZE? "<name>"
<name>
A string that represents the name of a measure configuration list
Details
This command returns the size (number of configuration indexes) of a measure configuration list. The
size of the list is equal to the number of configuration indexes in a configuration list.
Example
:SENSe:CONF:LIST:SIZE? "MyMeasList"
Returns the number of configuration indexes
in a measure configuration list named
MyMeasList
.
Example output:
3