2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-65
[:SENSe[1]]:CONFiguration:LIST:STORe
This command stores the active measure settings into the named configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Saved settings
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>"
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>", <index>
<name>
A string that represents the name of a measure configuration list
<index>
A number that defines a specific configuration index in the configuration list
Details
Use this command to store the active measure settings to a configuration index in a configuration list.
If the index parameter is not provided, the new settings are appended to the end of the list. The index
only stores the active settings for a single active measure function.
Configuration lists are not saved when the instrument is turned off or reset. To save a configuration
list, create a configuration script to save instrument settings, including any defined configuration lists.
Refer to
Instrument settings stored in a measure configuration list
(on page 4-86) for a complete list of
measure settings that the instrument stores.
Example
:SENSe:CONF:LIST:STOR "MyConfigList"
Stores the active settings of the instrument to
the end of the configuration list named
MyConfigList
.
:SENSe:CONF:LIST:STOR "MyConfigList", 5
Stores the active settings of the instrument to
the configuration list named
MyConfigList
in configuration index 5.
Also see
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 12-61)
(on page 12-9)