trigger.model.setblock() — trigger.BLOCK_CONFIG_PREV
This function defines a trigger model block that recalls the settings stored at the previous index in a source or
measure configuration list, or both a source and measure configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_CONFIG_PREV, "
configurationList
")
trigger.model.setblock(
blockNumber
, trigger.BLOCK_CONFIG_PREV, "
configurationList
",
"
optionalConfigList
")
blockNumber
The sequence of the block in the trigger model
configurationList
A string that defines the source or measure configuration list to
recall
optionalConfigList
A string that defines the second configuration list to recall the
index from; must be the opposite type of list than the first; for
example, if the first configuration list is a measure list, the second
configuration list must be a source list
Details
If two configuration lists are specified with this command, they must not be of the same type. For
example, if the first configuration list is a measure configuration list, the second configuration list must
be a source configuration list. The order of the configuration lists does not matter with this command,
as long as they are of the opposite type.
The Config List Prev block defines a trigger model block that recalls the settings stored at the
previous index in a source or measure configuration list if a single configuration list is specified. If
both measure and source configuration lists are specified, measure and source settings are each
recalled from the previous index in each list when this block is reached. The index numbers recalled
may not match; it depends on the number of indexes in each list and what index number each list is
on.
The configuration list previous index trigger block type recalls the previous index in a configuration
list. It configures the source or measure settings of the instrument based on the settings at that index.
The trigger model executes the settings at that index before the next block is executed.
The first time the trigger model encounters this block, the last index in the configuration list is recalled
if the list has not already had an index recalled by the recall block command in an earlier trigger
model block. If the configuration list has recalled an index with the recall block, the previous index in
the list is called instead of the last. For example, the recall block recalls index 1 by default, so if the
trigger model uses a recall block before this one, the first time the previous block is reached after that
recall, the last index is recalled. However, if the recall block recalled index 3, the previous block would
recall index 2. Each subsequent time trigger model execution reaches a configuration list previous
block for this configuration list, it goes backward one index. When the first index in the list is reached,
it goes to the last index in the configuration list.
The configuration lists must be defined before you can use this block.