Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-234
2470-901-01 Rev. A /
May
2019
Details
The branch-once building block branches to a specified block the first time trigger model execution
encounters the branch-once block. If it is encountered again, the trigger model ignores the block and
continues in the normal sequence.
The once block is reset when trigger model execution reaches the idle state. Therefore, the
branch-once block always executes the first time the trigger model execution encounters this block.
Example
trigger.model.setblock(2, trigger.BLOCK_BRANCH_ONCE, 4)
When the trigger model reaches block 2, the trigger model goes to block 4 instead of going in the default
sequence of block 3.
Also see
(on page 8-42)
trigger.model.setblock() —
trigger.BLOCK_BRANCH_ONCE_EXCLUDED
This function defines a trigger model block that causes the trigger model to go to a specified building block every
time the trigger model encounters it, except for the first time.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_BRANCH_ONCE_EXCLUDED,
branchToBlock
)
blockNumber
The sequence of the block in the trigger model
branchToBlock
The block number of the trigger model block to execute when the trigger model
encounters this block after the first encounter
Details
The branch-once-excluded block is ignored the first time the trigger model encounters it. After the first
encounter, the trigger model goes to the specified branching block.
The branch-once-excluded block is reset when the trigger model starts or is placed in idle.
Example
trigger.model.setblock(2, trigger.BLOCK_BRANCH_ONCE_EXCLUDED, 4)
When the trigger model reaches block 2 the first time, the trigger model goes to block 3. If the trigger model
reaches this block again, the trigger model goes to block 4.
Also see