Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-214
2470-901-01 Rev. A /
May
2019
If any events are detected before the trigger LAN connection is sent, the event is ignored and the
action overrun is set.
Trigger events
Event description
Event constant
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
A command interface trigger:
Any remote interface: *
TRG
GPIB only: GET bus command
USB only: A USBTMC TRIGGER message
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Digital input line edge (either rising, falling, or either based on
the configuration of the line) detected on digital input line
N
(1 to 6)
trigger.EVENT_DIGIO
N
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Appropriate LXI trigger packet is received on LAN trigger
object
N
(1 to 8)
trigger.EVENT_LAN
N
No trigger event
trigger.EVENT_NONE
Notify trigger block
N
(1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFY
N
Source limit condition occurs
trigger.EVENT_SOURCE_LIMIT
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Line edge detected on TSP-Link synchronization line
N
(1 to 3)
trigger.EVENT_TSPLINK
N
Example
trigger.lanout[5].stimulus = trigger.EVENT_TIMER1
Use the timer 1 trigger event as the source for LAN trigger 5 stimulus.
Also see
(on page 14-209)
trigger.model.abort()
This function stops all trigger model commands on the instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.model.abort()
Details
When this command is received, the instrument stops the trigger model.