Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-120
2470-901-01 Rev. A /
May
2019
Also see
(on page 14-46)
smu.measure.configlist.create()
(on page 14-115)
smu.measure.configlist.store()
This function stores the active measure settings into the named configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.measure.configlist.store("
listName
")
smu.measure.configlist.store("
listName
",
index
)
listName
A string that represents the name of a measure configuration list
index
A number that defines a specific configuration index in the configuration list
Details
Use this command to store the active measure settings to a configuration index in a configuration list.
If the index parameter is not provided, the new settings are appended to the end of the list. The index
only stores the active settings for a single active measure function.
Configuration lists are not saved when the instrument is turned off or reset. To save a configuration
list, create a configuration script to save instrument settings, including any defined configuration lists.
Refer to
Instrument settings stored in a measure configuration list
(on page 4-86) for a complete list of
measure settings that the instrument stores.
Example
smu.measure.configlist.store("MyConfigList")
Stores the active settings of the
instrument to the end of the configuration
list
MyConfigList
.
smu.measure.configlist.store("MyConfigList", 5)
Stores the active settings of the
instrument to configuration index 5 in the
measure configuration list
MyConfigList
.
Also see
(on page 14-46)
smu.measure.configlist.create()
(on page 14-115)