Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-122
2470-901-01 Rev. A /
May
2019
The options for
<bufferElements>
are described in the following table.
Option
Description
DATE
The date when the data point was measured
FORMatted
The measured value as it appears on the front panel
FRACtional
The fractional seconds for the data point when the data point was measured
READing
The measurement reading based on the
[:SENSe[1]]:FUNCtion[:ON]
setting; if
no buffer elements are defined, this option is used
RELative
The relative time when the data point was measured
SEConds
The seconds in UTC (Coordinated Universal Time) format when the data point was
measured
SOURce
The source value; if readback is ON, then it is the readback value, otherwise it is the
programmed source value (see
:SOURce[1]:<function>:READ:BACK
SOURFORMatted
The source value as it appears on the display
SOURSTATus
The status information associated with sourcing. The values returned indicate the
status of the following conditions:
Overvoltage protection was active
Measured source value was read
Overtemperature condition existed
Source function level was limited
Four-wire sense was used
Output was on
SOURUNIT
The unit of value associated with the source value
STATus
The status information associated with the measurement; see the "Buffer status bits
for sense measurements" table below
TIME
The time for the data point
TSTamp
The timestamp for the data point
UNIT
The unit of measure associated with the measurement
The
STATus
buffer element returns status values for the readings in the buffer. The status values are
integers that encode the status value. Refer to the following table for values.
Buffer status bits for sense measurements
Bit (hex) Name
Decimal Description
0x0001 STAT_QUESTIONABLE
1
Measure status questionable
0x0006 STAT_ORIGIN
6
A/D converter from which reading originated; for the
2470, this will always be 0 (main)
0x0008 STAT_TERMINAL
8
Measure terminal, front is 1, rear is 0
0x0010 STAT_LIMIT2_LOW
16
Measure status limit 2 low
0x0020 STAT_LIMIT2_HIGH
32
Measure status limit 2 high
0x0040 STAT_LIMIT1_LOW
64
Measure status limit 1 low
0x0080 STAT_LIMIT1_HIGH
128
Measure status limit 1 high
0x0100 STAT_START_GROUP
256
First reading in a group