![Tektronix Keithley SourceMeter 2470 Reference Manual Download Page 566](http://html1.mh-extra.com/html/tektronix/keithley-sourcemeter-2470/keithley-sourcemeter-2470_reference-manual_1078034566.webp)
2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-179
:TRIGger:LAN<n>:OUT:LOGic
This command sets the logic on which the trigger event detector and the output trigger generator operate on the
given trigger line.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
NEG
Usage
:TRIGger:LAN<n>:OUT:LOGic <logicType>
:TRIGger:LAN<n>:OUT:LOGic?
<n>
The LAN event number (1 to 8)
<logicType>
The type of logic:
POSitive
NEGative
Example
TRIG:LAN1:OUT:LOG POS
Set the logic to positive.
Also see
None
:TRIGger:LAN<n>:OUT:PROTocol
This command sets the LAN protocol to use for sending trigger messages.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
TCP
Usage
:TRIGger:LAN<n>:OUT:PROTocol <protocol>
:TRIGger:LAN<n>:OUT:PROTocol?
<n>
The LAN event number (1 to 8)
<protocol>
The protocol to use for messages from the trigger:
TCP
UDP
MULTicast