2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-185
status.questionable.condition
This attribute reads the Questionable Condition Register of the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
questionableRegister
= status.questionable.condition
questionableRegister
The value of the register (0 to 65535)
Details
This command reads the contents of the Questionable Condition Register, which is one of the
Questionable Event Registers.
For detail on interpreting the value of a register, see
(on page B-15).
Example
print(status.questionable.condition)
Reads the Questionable Condition Register.
Also see
(on page B-6)
(on page B-15)
status.questionable.enable
This attribute sets or reads the contents of the questionable event enable register of the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
status.preset()
Not applicable
0
Usage
questionableRegister
= status.questionable.enable
status.questionable.enable =
questionableRegister
questionableRegister
The value of the register (0 to 65535)
Details
This command sets or reads the contents of the Enable register of the Questionable Event Register.
When one of these bits is set, when the corresponding bit in the Questionable Event Register or
Questionable Condition Register is set, the MSB and QSM bits in the Status Byte Register are set.
For detail on interpreting the value of a register, see
(on page B-15).
Example
status.questionable.enable = 17
print(status.questionable.enable)
Set bits 0 and 4 of the Questionable Event Enable Register.
Returns 17, which indicates the register was set correctly.
Also see
(on page B-6)