2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-153
Example
print(smu.source.configlist.catalog())
Request the name of one source configuration list that
is stored in the instrument. Send the command again
until it returns
nil
to get all stored lists.
Also see
smu.source.configlist.create()
smu.source.configlist.create()
This function creates an empty source configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Configuration script
Usage
smu.source.configlist.create("
listName
")
listName
A string that represents the name of a source configuration list
Details
This command creates an empty configuration list. To add configuration indexes to this list, you need
to use the store command.
Configuration lists are not saved when the instrument is turned off. If you want to save a configuration
list through a power cycle, create a configuration script to save instrument settings, including any
defined configuration lists.
Example
reset()
smu.source.configlist.create("MyScrList")
print(smu.source.configlist.catalog())
print(smu.source.configlist.catalog())
smu.source.configlist.store("MyScrList")
smu.source.configlist.store("MyScrList")
print(smu.source.configlist.size("MyScrList"))
Create a source configuration list
named
MyScrLst
.
Print the name of one configuration list
stored in volatile memory.
Output:
MyScrList
Print the name of one configuration list.
Output:
nil
Nil indicates that no more configuration
lists are stored.
Store a configuration index in
MyScrList
.
Store a configuration index in
MyScrList
.
Print the number of configuration
indexes in
MyScrList
.
Output:
2