2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-15
If you send this command without the
<function>
parameter, it will set the constant for all measure
functions.
Example
CALC:VOLT:MATH:FORM PERC
CALC:VOLT:MATH:PERC 50
CALC:VOLT:MATH:STAT ON
Set the math operations for voltage to percent.
Set the percentage value to 50.
Enable math operations.
Also see
Calculations that you can apply to measurements
(on page 4-49)
:CALCulate[1]:<function>:MATH:FORMat
(on page 12-10)
:CALCulate[1]:<function>:MATH:STATe
(on page 12-15)
:CALCulate[1]:<function>:MATH:STATe
This command enables or disables math operation.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
OFF (0)
Usage
:CALCulate[1]:<function>:MATH:STATe <n>
:CALCulate[1]:<function>:MATH:STATe?
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<n>
Disable math operations:
OFF
or
0
Enable math operations:
ON
or
1
Details
When this command is set to on, the math operation specified by the math format command is
performed before completing a measurement.
If you send this command without the
<function>
parameter, it sets the state of math operations
for all measure functions.