2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-11
Math calculations are applied to the input signal after relative offset and before limit tests.
If you send this command without the
<function>
parameter, it will set the state of the math format
for all measure functions.
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STAT ON
Set the math function for voltage measurements to mx+b.
Set the scale factor for voltage measurements to 0.80.
Set the offset factor to 50.
Enable the math function.
Also see
Calculations that you can apply to measurements
(on page 4-49)
:CALCulate[1]:<function>:MATH:MBFactor
(on page 12-11)
:CALCulate[1]:<function>:MATH:MMFactor
:CALCulate[1]:<function>:MATH:PERCent
(on page 12-14)
:CALCulate[1]:<function>:MATH:STATe
(on page 12-15)
:CALCulate[1]:<function>:MATH:MBFactor
This command specifies the offset, b, for the y = mx + b operation.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0
Usage
:CALCulate[1]:<function>:MATH:MBFactor <n>
:CALCulate[1]:<function>:MATH:MBFactor <DEF|MIN|MAX>
:CALCulate[1]:<function>:MATH:MBFactor?
:CALCulate[1]:<function>:MATH:MBFactor? <DEF|MIN|MAX>
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<n>
The offset for the y = mx + b operation; the valid range is
−
1e12 to
+
1e12
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value