Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-138
2470-901-01 Rev. A /
May
2019
smu.measure.math.percent
This attribute specifies the reference constant that is used when math operations are set to percent.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
1
Usage
value
= smu.measure.math.percent
smu.measure.math.percent =
value
value
The reference used when the math operation is set to percent; the range is -1e12 to
+1e12
Details
This is the constant that is used when the math operation is set to percent for the selected
measurement function.
The percent math function displays measurements as percent deviation from a specified reference
constant. The percent calculation is:
Where:
Percent
is the result
Input
is the measurement (if relative offset is being used, this is the relative offset value)
Reference
is the user-specified constant
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.math.format = smu.MATH_PERCENT
smu.measure.math.percent = 50
smu.measure.math.enable = smu.ON
Set the measurement function to voltage.
Set the math operations to percent.
Set the percentage value to 50 for voltage
measurements.
Enable math operations.
Also see
Calculations that you can apply to measurements
(on page 4-49)
(on page 14-134)
(on page 14-135)