17
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
MAINTENANCE
3
9
Loosen the fastening clip of the exhaust gas line with a Phillips screwdriver.
10
Disconnect the power supply for the cell heater (plug connection).
.
Fig. 8: Power supply, cell heater
11
Disconnect the A-holder of the cell:
▸
Loosen the fastening screws on the left with a 5 mm Allen key (3 pieces).
▸
Loosen the fastening screws on the right with a 5 mm Allen key (3 pieces).
.
Fig. 9: Cell, A-holder, right
12
Lift the cell upwards and remove it from the cabinet.
The plug connection is located at slot CELL below the sender/receiver unit.
1 CELL slot
1 A-holder fastening screws, right