5
REPAIRS
8
Fit new mirror plate with new O-ring.
Fig. 108: O-ring, lower mirror plate
9
Fit the lower mirror plate.
Fig. 109: Fitting the lower mirror plate
10
Screw the lower mirror plate tight.
11
Screw the lower cell cover plate tight.
12
Position the tubes sealing rings.
13
Refit the cell (
see Chapter “Fitting the complete cell”, page 18
14
Connect the MCS200HW power supply.
15
Switch the MCS200HW on.
16
Allow the MCS200HW to heat up (for at least 2 hours).
17
Perform zero point adjustment (
see Chapter “Adjusting the zero point”, page 128
).
18
Check reference energy (
see Chapter “Reference energy check”, page 47
1 Lower mirror plate
2 O-ring
NOTE: Material damage
Protect mirror plates from insulation particles. Do not touch the insulation material.
Observe alignment: Align the labeled side of the mirror plate to the front of the cell inlet
filter.
1 Mirror plate labeling
2 Cell front side with cell inlet filter