8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
Fig. 183: Filter wheel housing 3, fastening screw, left
3
Remove the filter wheel housing.
Fig. 184: LPMS03 detector circuit board
4
Disconnect the detector cable.
5
Loosen the fastening screws of the LPMS03 detector circuit board with a Phillips screw-
driver (4 pieces).
6
Screw the new LPMS03 detector circuit board tight.
7
Connect the detector cable to the LPMS03 detector circuit board.
8
Insert filter wheel housing 3 into the sender/receiver unit and screw tight.
▸
The filter wheel housing cover points upwards.
9
Connect the lug connection of the filter wheel housing to the LPMS04 circuit board.
10
Close the sender/receiver unit cover and screw tight.
11
Connect the MCS200HW power supply.
12
Perform optical adjustment (
see Chapter “Performing optical adjustment”, page 103
).
13
Reset the reference energy (
see Chapter “Resetting the reference energy”, page 91
).
14
Perform zero point adjustment (
see Chapter “Adjusting the zero point”, page 128
1 Fastening screw
1 Fastening screw
2 LPMS03 detector circuit board