8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
9
Insert filter wheel housing 1 into the sender/receiver unit and screw tight.
10
Connect the lug connection of filter wheel housing 1 to the LPMS04 circuit board.
11
Replace the drying agent (
see Chapter “Replacing the drying agent”, page 44
).
12
Close the sender/receiver unit cover and screw tight.
13
Connect the MCS200HW power supply.
14
Perform optical adjustment (
see Chapter “Performing optical adjustment”, page 103
).
5.3.12
Replacing the LPMS03 detector cable
Fig. 179: LPMS03 Detector cable
1
Disconnect the MCS200HW power supply.
2
Open the sender/receiver unit (
see Chapter “Opening the sender/receiver unit”,
).
3
Disconnect the detector cable from the LPMS04 detector circuit board and LPMS03
detector circuit board.
4
Connect a new detector cable.
Fig. 180: Detector cable
5
Replace the drying agent (
see Chapter “Replacing the drying agent”, page 44
).
6
Close the sender/receiver unit cover and screw tight.
7
Connect the MCS200HW power supply.
Spare parts set No.
Contents
2099806
Detector cable
1
Detector cable
The drying agent must be replaced each time the sender/receiver unit is opened.
Connectors on both ends of the detector cable are identical.