79
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
5.2.5
Replacing the heating cartridges
To replace the heating cartridges, the complete cell must be removed and all covers
removed.
Fig. 110: Heating cartridges
1
Remove the complete cell (
see Chapter “Removing the complete cell”, page 65
2
Place the cell down with the rear side upwards.
>>> The side with the cell inlet filter and ejector is at the bottom.
Fig. 111: Cell, overview, left
3
Loosen the fastening clips of the heating cartridges with a Phillips screwdriver (9 pieces).
Tools required
Phillips screwdriver
Slot screwdriver
Pliers
Spare parts set No.
Contents
2099812
Heating cartridge garlands set, 2-HP 115 V, 180 W, 10 x 80
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.
1
Heating cartridges
2
Grounding
3
Cable to thermostatic switch
4
PT100
Document the arrangement of the cables of the heating cartridges for reassembly, e.g.
take a photo.