101
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
6
Loosen the plug connection of the pressure control module.
7
Disconnect the plug connection of the pressure control module.
8
Loosen the fastening screws of the pressure control module with a 5 mm Allen key
(2 pieces).
Fig. 149: Pressure control module, fastening screws
9
Remove the pressure control module.
10
Screw the new pressure control module tight.
11
Connect the plug connection.
12
Connect the hose connection of the pressure control module.
13
Connect the hoses to the slots.
Fig. 150: Hose slots
14
Replace the electronics housing cover and screw tight.
With the cell cover closed, mark the hoses before removing them.
1
Fastening screw
Sequence (if the hose connections have not been marked):
– Green, rear
– Blue (instrument air), front
– From top to bottom on the cell corresponds to from back to front on pressure control module