35
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
MAINTENANCE
3
.
Fig. 39: Fastening screws, upper mirror plate
6
Lever out the mirror plate:
1
Screw the screws into the thread openings provided on the mirror plate (2 pieces).
Fig. 40: Thread openings, upper mirror plate
2
Tighten the fastening screws with a 5 mm Allen key.
▸
Tighten fastening screws parallel.
▸
Ensure the screws grip properly and avoid thread seizure.
7
Remove the upper mirror plate.
1 Fastening screws
Keep screws, spring washers and washers for reassembly.
1 Thread openings
Alternative:
– Unscrew and remove all cell covers (
see Chapter “Removing the complete cell”,
– Apply the flat side of a screwdriver to the side recesses of the mirror plate.
– Lever out the mirror plate.
NOTE: Material damage
Do not touch the coated surfaces of the mirror plates.