REPAIRS
5
8
Place the upper electronics unit with the new LPMS01 on the lower electronics unit.
9
Connect the cables to the lower electronics unit.
10
Screw the upper and lower electronics unit together:
▸
Top fastening screws (2 pieces).
▸
Rear fastening screws (2 pieces).
11
Refit the CPU (
see Chapter “Replacing the LINUX CPU”, page 135
12
Connect the connection cable to the LPMS04.
13
Insert the electronics unit into the device cabinet and screw tight (
ing and fitting the electronics unit”, page 129
14
Perform zero point adjustment (
see Chapter “Adjusting the zero point”, page 128
5.4.3
Replacing the LINUX CPU
Fig. 218: Linux CPU
1
Remove the electronics unit (
see Chapter “Removing and fitting the electronics unit”,
).
2
Loosen the CPU housing fastening screws with a Phillips screwdriver (5 pieces).
Fig. 220: Linux CPU housing
3
Remove the CPU housing.
4
Loosen the studs with a 5.5 mm socket wrench (5 pieces).
Tools required
Phillips screwdriver
5.5 mm socket wrench
Spare parts set No.
Contents
2085767
LINUX CPU
Create a parameter backup with SOPAS when possible:
1
Select
Maintenance
-> ‚
Load/Save Parameter
.
1
Click
Save parameter backup
.
Replace the SD card when a parameter backup cannot be created.
1
Screws