REPAIRS
5
5.3.3
Removing the complete sender/receiver unit
1
Remove the cell (
see Chapter “Removing the complete cell”, page 15
).
2
Remove the electronics housing cover (
see Chapter “Removing the electronics housing
3
Disconnect all plug connections.
4
Open the sender/receiver unit (
see Chapter “Opening the sender/receiver unit”,
).
5
Loosen the fastening screws inside with a 6 mm Allen key.
Fig. 134: Sender/receiver unit, fastening screws, inside, right
Tools required
Phillips screwdriver
6 mm Allen key
2.5 mm Allen key
Tweezers
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.
The drying agent must be replaced each time the sender/receiver unit is opened.
1
Fastening screw