5
REPAIRS
1
Adjust adjusting screws on the beamer and receiver (2 pieces each).
Fig. 154: Sender/receiver unit, adjusting screws
2
Check adjustment (two options):
a) SOPAS display
▸
Select
Diagnosis
->
Hardware
.
▸
Adjust
VVM Magnitude
to the maximum possible value.
Fig. 155: SOPAS VVM Magnitude
b) LED on LPMS04
>>> Green LED shows increasing energy.
>>> Red LED shows decreasing energy.
4
Click
Test measurement
to deactivate.
5
Replace the drying agent (
see Chapter “Replacing the drying agent”, page 44
6
Close the sender/receiver unit cover and screw tight.
1
Adjusting screws
2
Adjusting screws
3
LED
The SOPAS display updates only every 3 to 4 seconds.