103
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
1
Disconnect the MCS200HW power supply.
2
Open the sender/receiver unit (
see Chapter “Opening the sender/receiver unit”,
)
3
Disconnect the plug connection of the power supply.
4
Disconnect the motor control plug connection.
5
Loosen the fastening screws with a 5 mm Allen key (2 pieces).
6
Screws remain on the beamer.
7
Remove the complete beamer unit.
8
Screw the new beamer unit tight.
9
Connect the plug connections.
10
Replace the drying agent (
see Chapter “Replacing the drying agent”, page 44
).
11
Perform optical adjustment (
see Chapter “Performing optical adjustment”, page 103
).
12
Close the sender/receiver unit cover and screw tight.
13
Connect the MCS200HW power supply.
14
Reset the beamer service life (
see Chapter “Resetting the beamer service life”,
).
5.3.8
Performing optical adjustment
1
Start the test measurement to stop the rotating cycle of the filter wheels.
1
Select
Global definitions
.
2
Click
Test measuring
.
Fig. 153: Performing test measurement
2
Open the sender/receiver unit (
see Chapter “Opening the sender/receiver unit”,
).
3
Maximize energy on the beamer and receiver.
The drying agent must be replaced each time the sender/receiver unit is opened.
Pay attention to dust protection: Replace the electronics housing cover during work on
the sender/receiver unit.
The drying agent must be replaced each time the sender/receiver unit is opened.