R&S FSMR
DISPlay Subsystem
1313.9723.12
6.1-101
E-1
DISPlay[:WINDow<1|2>]:ACTive?
This command returns the active measurement window. The numeric response has following
meaning:
1
Screen A
2
Screen B
3
Screen C
4
Screen D
Example:
"DISP:WIND:ACT?
"
returns the active window
Characteristics:
*RST value:
--
SCPI:
device-specific
DISPlay[:WINDow<1|2>]:SELect
This command selects the active measurement window. WINDow1 corresponds to SCREEN A,
WINDow2 to SCREEN B.
In FULL SCREEN mode, the measurements are only performed in the active measurement window.
Measurements are therefore initiated in the active window and result queries (marker, trace data and
other results) answered also in the active window.
Initiating measurements and queryíng results in the inactive window yields an error message
(execution error).
In split screen mode, the selection of the active window for result queries is irrelevant.
In FULL SCREEN mode, settings
can also be performed in the inactive measurement window. They
become effective as soon as the corresponding window becomes active.
This command is not available in GSM/EDGE mode. In GSM/EDGE mode, the display
is always set to FULL SCREEN, which corresponds to SCREEN A and thus to
WINDow1
.
WINDow1
is automatically selected when the GSM/EDGE mode is
activated with command
INSTrument:SELect:MGSM.
Example:
"
DISP:WIND2:SEL
'Selects SCREEN B as active measurement window.
Characteristics:
*RST value:
SCREEN A active
SCPI:
device-specific
This command is an event and therefore has no query.
DISPlay[:WINDow<1|2>]:SIZE LARGe | SMALl
This command switches the measurement window for channel and adjacent-channel power measurements
or for active FM demodulator (FS-K7) to full screen or half screen. Only "1" is allowed as a numerical
suffix.
Example:
"DISP:WIND1:SIZE LARG"
'Switches the measurement window to
full screen.
Characteristics:
*RST value:
SMALl
SCPI:
device-specific
DISPlay[:WINDow<1|2>]:TEXT[:DATA]
<string>
This command defines a comment (max. 20 characters) which can be displayed on the screen in
the selected measurement window.
Example:
"
DISP:WIND2:TEXT 'Noise Measurement'"
'Defines the title for screen B
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...