SENSe:MPOWer Subsystem
R&S FSMR
1313.9723.12
6.1-178
E-1
COMMAND
PARAMETER
UNIT
COMMENT
[SENSe<1|2>]
:MPOWer
:RESult
[:LIST]?
:MIN?
Query only
Query only
[:SEQuence]
<numeric_value>,
<numeric_value>,
<numeric_value>,
EXTernal | VIDeo,
<numeric_value>,
<numeric_value>,
MEAN | PEAK,
<numeric_value>
HZ,
HZ,
S,
--,
PCT,
S,
--,
--
[SENSe:MPOWer:RESult[:LIST]?
This command queries the results of a multiple burst power measurement as configured and initiated
with
SENSe:MPOWer[:SEQuence]
. The results are output in a comma-separated list of floating
point values. The unit used for the return values is always dBm.
This command may be used to obtain measurement results in an asynchronous way using the
service request mechanism for synchronization with the end of the measurement.
If no measurement results are available, the command will return a query error.
This command is not available for GSM/EDGE measurements modulation accuracy
(MAC), phase/frequency error (PFE) and power versus time (PVT).
Example:
*ESE 1
'Configuration of status reporting systems for the
*SRE 32
'generation of an SRQ on operation complete
SENSe:MPOWer
935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
'Configuring and starting the measurement
...
'Further actions of the control computer during
'measurement
On SRQ:
'Response to service request
SENSe:MPOWer:RESult?
Characteristics
: *RST value:
--
SCPI:
instrument-specific
SENSe:MPOWer:RESult:MIN?
This command queries the minimum power value in a multiple burst power measurement as configured
and initiated with
SENSe:MPOWer[:SEQuence]
. The unit used for the return values is always dBm.
If no measurement result is available, the command will return a query error.
This command is not available for GSM/EDGE measurements modulation accuracy
(MAC), phase/frequency error (PFE) and power versus time (PVT).
Example:
*ESE 1
'Configuration of status reporting systems for the
*SRE 32
'generation of an SRQ on operation complete
SENSe:MPOWer
935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
'Configuring and starting the measurement
...
'Further actions of the control computer during
'measurement
On SRQ:
'Response to service request
SENSe:MPOWer:RESult:MIN?
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...