R&S FSMR
MKR – Analyzer
1166.3363.12
4.13-7
E-1
LINK MKR1
AND DELTA1
With the softkey
LINK MKR1 AND DELTA1
the delta marker1 can be linked to
marker1, so if the x-axis value of the marker1 is changed the delta marker1 will
follow on the same x-position. The link is default off, and can be switched on.
Example for setup:
•
PRESET
•
TRACE | MAX HOLD
•
TRACE | SELECT TRACE | 2 | AVERAGE
•
MKR (Switches marker1 on)
•
MARKER NORM DELTA | DELTA (Delta Marker 1 ON)
•
MKR-> | MKR->TRACE | 2
•
LINK MKR1 AND DELTA1
Now select the Marker1 (by switching MARKER1 from DELTA back to NORM)
and when changing the x-axis value (by knob wheel or UP/DOWN keys) the
delta marker1 will follow automatically.
The delta marker1 x-value can not be changed away from 0 as long as the link
functionality is active.
IEC/IEEE-bus command:
CALC1:DELT1:LINK ON | OFF
MARKER
CNT RESOL
10 kHz
CNT RESOL
1 kHz
CNT RESOL
100 Hz
CNT RESOL
10 Hz
CNT RESOL
1 Hz
CNT RESOL
0.1 Hz
The
CNT RESOL ...
softkeys select the counter resolution. They are selection
switches, i.e. only one of the can be active at any one time.
The marker stop time, ie the frequency measurement time, depends on the
selected resolution.
IEC/IEEE-bus command:
CALC:MARK1:COUN:RES <value>
Measurement example:
The frequency of a CW signal is to be determined by means of the frequency
counter with a resolution of 10 Hz.
[
PRESET
]
R&S FSMR is set to the default setting.
[
MARKER
]
MARKER 1
is switched on and set to the maximum value
of the displayed spectrum.
[
SIGNAL
COUNT
]
The frequency counter is switched on. R&S FSMR counts
the frequency of the signal at the marker position with a
resolution of 1 kHz. The counted frequency is indicated in
the marker info field.
[
NEXT
]
Changes to the submenu for setting the counter resolution.
[
CNT RESOL
10 Hz
]
The frequency counter resolution is increased to 10 Hz.
STEPSIZE
STANDARD
STEPSIZE
SWP POINTS
These softkeys control the knob increment/decrement of the marker position.
STEPSIZE STANDARD uses the grid resolution (span/625), STEPSIZE SWP
POINTS uses the available sweep points configured in the SWEEP menu.
IEC/IEEE-bus command:
CALC1:MARK:X:SSIZ STAN | POIN
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...