R&S FSMR
SENSe:CORRection Subsystem
1313.9723.12
6.1-155
E-1
SENSe:CORRection Subsystem
The SENSe:CORRection subsystem controls the correction of measured results by means of
frequency-dependent correction factors (e. g. for antenna or cable attenuation).
This subsystem also controls calibration and normalization during operation with the external generator control
option (B10). The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).
The commands of this subsystem are not available during GSM measurements.
COMMAND
PARAMETERS
UNIT
COMMENT
[SENSe<1|2>]
:CORRection
:COLLect
[:ACQuire]
INPut| PMeter | PSPLitterTHRough | OPEN
no query
:METHod
TRANsmission | REFLexion
:PLOSs
:INPut
:SPATh
:STATe
<freq>,<level>..
<numeric_value>
<Boolean>
DB
:RECalll
no query
[:STATe]
<Boolean>
:TRANsducer
:ACTive?
query only
:ADJust
:RLEVel
[:STATe]
<Boolean>
:CATalog?
query only
:COMMent
<string>
:DATA
<freq> , <level> ...
HZ
:DELete
no query
:GENerate
<name>
:SCALing
LINear|LOGarithmic
:SELect
<name>
[:STATe]
<Boolean>
:UNIT
<string>
:VIEW
<Boolean>
:VSWR
[:STATe]
ON|OFF
:YIG
:TEMPerature
[:AUTO]
<Boolean>
[SENSe<1|2>:]CORRection:COLLect[:ACQuire]
INPut | PMETer | PSPLitter | THRough | OPEN
This command carries out the various steps of absolute level calibration in the
Tuned RF Level
mode.
PMETer
Measures the reference level with a power meter.
INPut
Measures the comparison value at the receiver's RF input and corrects it to the
reference level.
PSPLitter
Measures the reference level and then the comparison value.
For this measurement, the power meter and the receiver's RF input must be
connected to the DUT via a power splitter.
THRough
"TRANsmission" mode:
calibration with direct connection between tracking
generator and device input.
"REFLection" mode:
calibration with short circuit at the input
OPEN
only valid in "REFLection" mode: calibration with open input
Example:
"CORR:COLL PSPL;*WAI"
'Starts the reference measurement and the
comparison measurement one after the other
and waits for the operation to be completed.
Characteristics:
*RST value:
--
SCPI:
device-specific
This command is an event and therefore has no *RST value and no query.
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...