R&S FSMR
SENSe:ADEMod - Subsystem
1313.9723.12
6.1-139
E-1
[SENSe:]ADEMod:DETector
PPEak | MPEak | PAVerage | RMS | AVERage | SRMS | THD | SINad
[[,PPEak | MPEak | PAVerage | RMS | AVERage | SRMS | THD | SINad],…]
This command switches on the detectors whose results are to be displayed in the modulation
summary.
PPEak
Result of measurement with +PK detector
MPEak
Result of measurement with -PK detector
PAVerage
Result of averaging
±
PK/2, i.e. (plus peak + minus peak)/2
AVERage
Result of measurement with average detector
RMS
Result of measurement with RMS detector
SRMS
Result of measurement with
\
2 RMS detector
THD
Total harmonic distortion
SINad
Signal, noise and distortion
THD and SINad are always displayed together. The above command can be used to
activate any combination of detectors. For the detailed configuration of the individual
detectors, including relative measurements, see the commands below.
Example:
"ADEM:DET PPE, RMS, THD"
'The peak and RMS detectors are switched on
together with the THD and the SINAD display.
Characteristics:
*RST value:
PPE
SCPI:
device-specific
[SENSe:]ADEMod:DETector:AVERage:MODE
ABSolute | RELative
This command switches on absolute or relative measurement for the average detector in the
modulation summary. Each time the relative measurement mode is switched on, the current
measured value is taken as a reference for the subsequent relative measurements. The command is
executed independently of the settings for the other detectors.
Example:
"
ADEM:DET:AVER:MODE REL
"
'Switches on the relative measurement
'for the average detector.
Characteristics:
*RST value:
ABS
SCPI:
device-specific
[SENSe:]ADEMod:DETector:AVERage:REFerence
<numeric_value>
This command sets or queries the reference value for the relative measurement for the average
detector in the modulation summary.
Example:
"
ADEM ON
"
'Switches on the demodulator.
"
CALC2:FEED 'XTIM:FM:TDOM'
" 'Switches on the FM modulation summary.
"
ADEM:DET:AVER ON
"
'Switches on the average detector.
"
ADEM:DET:AVER:MODE REL
"
'Switches on the relative measurement
for the average detector.
"
ADEM:DET:AVER:REF 20 kHz
" 'Sets the reference value.
Characteristics:
*RST value:
0
SCPI:
device-specific
[SENSe:]ADEMod:DETector:AVERage:REFerence:AUTO
ONCE
This command uses the current measured value as a reference for the relative measurement for the
average detector in the modulation summary.
Example:
"
ADEM:DET:AVER:REF:AUTO ONCE
" 'Uses the current measured value
'as a reference for the relative
'measurement.
Characteristics:
*RST value
-
SCPI:
device-specific
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...