CALCulate Subsystem
R&S FSMR
1313.9723.12
6.1-52
E-1
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:MDEPth:RESult?
This command queries the AM modulation depth in the indicated measurement window.
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single
sweep mode.
The numeric suffix <1 to 4> of :MARKer is irrelevant for this command.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:X 10MHZ"
'Sets the reference marker (marker 1) to
'the carrier signal at 10 MHz.
"CALC:MARK:FUNC:MDEP ON"
'Switches on the modulation depth
'measurement in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:MDEP:RES?"
'Outputs the measured value of screen A.
Characteristics:
*RST value:
-
SCPI:
device-specific
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:MDEPth:[:STATe]
This command switches on the measurement of the AM modulation depth. An AM-modulated carrier
is required on the screen for correct operation. If necessary, marker 1 is previously activated and set
to the largest signal available.
The level value of marker 1 is regarded as the carrier level. On activating the function, marker 2 and
marker 3 are automatically set as delta markers symetrically to the carrier to the adjacent maxima of
the trace.
If the position of delta marker 2 is changed, delta marker 3 is moved symetrically with respect to the
reference marker (marker 1). If the position of delta marker 3 is changed, fine adjustment can be
performed independently of delta marker 2.
The R&S FSMR calculates the power at the marker positions from the measured levels.
The AM modulation depth is calculated from the ratio of power values at the reference marker and
the delta markers. If the two AM sidebands differ in power, the average value of the two power
values is used for calculating the AM modulation depth.
The numeric suffix <1 to 4> of :MARKer is irrelevant with this command.
Example:
"CALC:MARK:X 10MHZ"
'Sets the reference marker (marker 1) to
'the carrier signal at 10 MHz
"
CALC:MARK:FUNC:MDEP ON"
'Switches on the modulation depth
'measurement in screen A.
"CALC:DELT2:X 10KHZ"
'Sets delta markers 2 and 3 to the signals
'at 10 kHz from the carrier signal
"CALC:DELT3:X 9.999KHZ"
'
Corrects the position of delta marker 3
'relative to delta marker 2.
Characteristics:
*RST value:
OFF
SCPI:
device-specific
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...