DIAGnostic Subsystem
R&S FSMR
1313.9723.12
6.1-96
E-1
DIAGnostic<1|2>:SERVice:INPut:PULSed:PRATe
10 kHz | 62.5 kHz | 1 MHz | 128 MHz | 640 MHz
This command selects the pulse rate for the pulsed calibration signal in the selected measurement window.
Available pulse frequencies are 10 kHz, 62.5 kHz, 100 kHz, 1 MHz, 128 MHz and 640 MHz.
Example:
"DIAG:SERV:INP:PRAT 128 MHz"
Characteristics:
*RST value:
128 MHz
SCPI:
device-specific
DIAGnostic<1|2>:SERVice:INPut:PULSed[:STATe]
ON | OFF
This command toggles the calibration signal in the selected measurement window between pulsed
and non-pulsed. The selection takes effect only if the RF input has been set to the internal reference
signal using
DIAG:SERV:INP CAL
.
Example:
"DIAG:SERV:INP CAL;
DIAG:SERV:INP:PULS ON"
Characteristics:
*RST value:
OFF
SCPI:
device-specific
DIAGnostic<1|2>:SERVice:INPut:RECTangle:PRATe
<numeric_value>
This command sets the frequency of the rectangle calibration signal in the selected measurement
window (suffix <1|2>).
5 kHz | 31.25 kHz | 50 kHz | 250 kHz | 500 kHz can be set.
Example:
"DIAG:SERV:INP:RECT:PRAT 128 MHz"
Characteristics:
*RST value:
5 kHz
SCPI:
device-specific
DIAGnostic<1|2>:SERVice:INPut[:SELect]
CALibration | RF
This command toggles between the RF input on the front panel and the internal 128-MHz reference
signal in the selected measurement window. The level of the 128-MHz signals can be selected by
command
DIAG:SERV:CSOurce
.
Example:
"DIAG:SERV:INP CAL"
Characteristics:
*RST value:
RF
SCPI:
device-specific
DIAGnostic<1|2>:SERVice:INPut:RECTangle[:STATe]
ON | OFF
This command switches the rectangle calibration signal on or off in the selected measurement window
(suffix <1|2>). The setting takes effect only if the RF input is set to the internal reference signal
(
DIAG:SERV:INP CAL
command).
If the rectangle calibration signal is switched on, the pulsed calibration signal is
switched off.
Example:
"DIAG:SERV:INP CAL;
DIAG:SERV:INP:RECT ON"
Characteristics:
*RST value:
OFF
SCPI:
device-specific
Summary of Contents for FSMR
Page 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Page 4: ......
Page 22: ......
Page 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Page 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Page 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Page 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Page 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Page 66: ......
Page 86: ......
Page 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Page 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Page 144: ......
Page 150: ......
Page 156: ......
Page 202: ......
Page 208: ......
Page 216: ......
Page 230: ......
Page 254: ......
Page 280: ......
Page 348: ......
Page 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Page 390: ......
Page 408: ......
Page 428: ......
Page 446: ......
Page 536: ......
Page 799: ......
Page 921: ......
Page 935: ......
Page 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...