Calibration Start-Up
It is possible to start calibration in two manners:
• Press the LIGHT (C) push button and keep
pressed while switching the collimator ON,
see
Fig. 1.
• jumper J4 on the external board, see
Fig.
1
. The systems sees the insertion and
displays the following:
Remove Jumper to start
Fig. 2
By removing the jumper, the procedure starts with the calibration of the SID potentiometers.
Calibration Procedure
1. Language Selection
The display will read one of the two following messages, depending on the current language
setting.
A
D
C
B
A - Encoder Cross changes option and set up the
Cross cassette size.
B - Encoder Long sets up Long cassette size.
C - to confirm
D - to skip without saving
Fig. 1
Language
English