Model R 221 ACS - R 221 ACS DHHS
MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
C
AN
B
US
M
ESSAGES
71
MTR
2
21
AC
S
-
MTR 22
1
ACS DHHS - Can Bus Me
ssages
Tx_Con
fig
0x600
8
0
0x4B
EnableManCollimation
0x00
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator manual collimation with the
provided setting value.
EnableManCollimation = 0x00 : the X-Ray Collimator manual
collimation disabled.
EnableManCollimation = 0x01 : the X-Ray Collimator manual
collimation is enabled. (default)
1.07
Tx_Con
fig
0x600
8
0
0x4D
FilterBtnEnabing
0x00
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator Filter Button installed on the
Front Panel.
FilterBtnEnabing = 0x00 : the X-Ray Collimator front panel filter
button is disabled.
FilterBtnEnabing = 0x01 : the X-Ray Collimator front panel filter
button is enabled.
1.07
Tx_Con
fig
0x600
8
0
0x4E
LightBtnEnabing
0x00
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator Light Button installed on the
Front Panel.
LightBtnEnabing = 0x00 : the X-Ray Collimator front panel Light
button is disabled.
LightBtnEnabing = 0x01 : the X-Ray Collimator front panel Light
button is enabled.
1.07
Tx_Con
fig
0x600
8
0
0x4F
lightTrigTime
0x00
0x00
0x00
0x00
0x00
0x00
Configure the Light Trigger time when shutters are moved by
using knobs.
lightTrigTime [5;60] s: defaut : 6s.
1.09
Tx_Con
fig
0x600
8
0
0x50
LSB_MaxFixedLeftSIDValue
MSB_MaxFix
edLeftSIDValu
e
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator maximum fixed Left SID (Source
Image Distance) with the provided setting value defined with
bytes [MSB_MaxFixedLeftSIDValue;
LSB_MaxMaxFixedLeftSIDValue].
MaxFixedLeftSIDValue : 100 cm (default)
MaxFixedLeftSIDValue eligible range of values : [50;300]cm
1.10
Tx_Con
fig
0x600
8
0
0x51
LSB_MaxFixedRightSIDValue
MSB_MaxFix
edRightSIDVa
lue
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator maximum fixed Right SID
(Source Image Distance) with the provided setting value defined
with bytes [MSB_MaxFixedRightSIDValue;
LSB_MaxFixedRightSIDValue].
MaxFixedRightSIDValue : 100 cm (default)
MaxFixedRightSIDValue eligible range of values : [50;300]cm
1.10
Tx_Con
fig
0x600
8
0
0x52
MSB_0x7F5
LSB_0x7F5
0x00
0x00
0x00
0x00
0x00
Command message used to set the 0x7F5 output frequency
status messages (in
msec
). (default value : 0ms)
MSB_0x7F5, LSB_0x7F5 : output frequency for the 0x7F5
status message.
MSB_0x7F5, LSB_0x7F5 eligible range of values : [50;1000] ms
Note : if the frequency output message set is equal to 0ms, the
output message is not emitted .
1.13
Type
ID
DLC RTR
D0
D1
D2
D3
D4
D5
D6
D7
Description
Fro
m
FW
Ver
sion